TM510FNplus Coating thickness gauge
● 2 Measuring modes: continuous/single
● 2 Shutdown modes: manual/automatic
●Wide measuring range with 5 probes available (next page)
●Direct testing mode and block statistics mode (APPL/BATCH)
●Can connect with printer to out of statistical values
●Dataview to connect with PC with USB 2.0 port
●500 datas can be stored
Main features
Can use various probe (F400, F1, F1/90 °, F10, N1, N400, etc.) measurement;
Three calibration methods: one point calibration, two point calibration, the basic calibration;
Display resolution: 0.1 um (measuring range of less than 100 um)
1μ m (range greater than 100 u m)
Have five statistics, data storage 500
There are two working methods: direct ways and means of group
There are two measurements: continuous measurement and a single measurement
There are two shutdown: manual and automatic shutdown shutdown
Can be set Bound: The gauge of the measured value can be automatic alarm and a number of measurements available on the histogram value analysis;
Deleted features: the gross error and error settings can be deleted;
Printing: Print Measurement measurement, statistics, gauges, histogram
A music tone in the operation carried out at any time tips
A power supply under-voltage direct function
An error function
And printers, computer communications (communications software operating environment for the Window operating system) connectivity.
Technical Specifications:
Measuring range | 0-1250µ with standard probe F1.N1 (10.000mm max) |
Probes available | 5 probes available for F (ferrous: on steel/iron) and N (non-ferrous metals) |
Tolerance | F1:±(1µ+3%H) N1.±(1.5µ+3% H)H: actual thickness tested |
Resolution | Alphanumeric with 4 large digits |
Operation language | English |
Standards | DIN, ISO, ASTM,BS |
Min. measuring area | F1:(standard probe) |
Min. curvature radius | convex:3mm, concave:50mm |
Min. substratethickness | type F: 0.5mm, type N: 50mm |
Calibration | Zero and foil calibration |
Statistics | Number of measurements, mean, standard deviation,maximum and minimum of 3000readings |
Data memory | 500 measuring data |
Limits | Adjustable with acoustic alarm |
Interface | USB 2.0 |
Operating temperature | 0-40℃ |
power supply | AA size 1.5V |
Dimensions | 125*67*31mm (main unit) |
Weight | 345.g |
Standard Delivery
· Main unit | 1 |
· Probe F1 or N1 | 1 |
· Calibration foils | 5 |
· Instruction manual | 1 |
· Cetificate | 1 |
· Warranty card | 1 |
· Carrying case | 1 |
Communication cable | 1 |
PC software Dataview | 1 |
Optional accessories
●5 probes for different applications
●Calibration foils in various thickness
Type F probe:
Probe |
F400 |
F1 |
F1/90° |
F10 |
|||
Measuring Principle |
Magnetic method |
||||||
Measuring range(mm) |
0~400um |
0~1250 um |
0~10000 |
||||
Min resolution(mm) |
0.1 |
0.1 |
10 |
||||
Tolerance |
One-point calibration (mm) |
±(2%H+0.7) |
±(2%H+1) |
±(2%H+10) |
|||
Two-point calibration (mm) |
±(1%H+0.7) |
±((1%H+1) |
±(1%H+10) |
||||
Minimum radius of curvature (convex) (mm) |
1 |
1.5 |
flat |
10 |
|||
Minimum measuring area(mm) |
F3 |
F7 |
F7 |
F40 |
|||
Minimum thickness of base material(mm) |
0.2 |
0.5 |
0.5 |
2 |
Type N probe
Probe | N400 |
N1 |
CN02 |
|
Measuring Principle |
Eddy current method |
|||
Measuring range (mm) |
0~400 |
0~1250 |
10~200 |
|
Min resolution(mm) |
0.1 |
0.1 |
1 |
|
Tolerance | One-point calibration (mm) |
±(2%H+0.7) |
±(2%H+1.5) |
±(2%H+1) |
Two-point calibration (mm) | ±(1%H+0.7) | ±(1%H+1.5) |
-------- |
|
Minimum radius of curvature (convex) (mm) |
1.5 |
3 |
flat |
|
Minimum measuring area(mm) |
F4 |
F5 |
F7 |
|
Minimum thickness of base material(mm) |
0.3 |
0.3 |
0 |