Have a question? Give us a call: +8613911515082

TMHR-150SX Rockwell hardness standard machine

Short Description:


Product Detail

Product Tags

I、Product Introduction

TMHR-150SX Rockwell hardness standard machine is a new generation of Rockwell hardness standard machine with international advanced level; Mainly suitable for the use of standard Rockwell hardness block verification, Rockwell hardness measurement value transfer.

The standard machine has the characteristics of high automation, high precision, good stability, etc., and fully meets the index requirements of the Rockwell hardness standard machine in China and the United States.

 

  1. Rockwell A ~ K scale;
  2. Automatic closed-loop load control, one-key operation, automatic loading, automatic unloading and indentation depth measurement;
  3. Automatic application of initial test force and main test force, automatic removal of main test force;
  4. The maximum allowable deviation of the initial test force is less than ±0.2%, and the maximum allowable deviation of the total test force is less than ±0.1%;
  5. The resolution of the indentation depth measurement system is 0.1μm, and the maximum allowable error is less than ±0.2μm;
  6. It can be set and controlled strictly during each period of time during loading and unloading.
  7. The uncertainty of test time is less than 0.5 seconds;
  8. The Angle error of diamond compression head Angle is less than ± 0.1
  9. Diameter error of ball indenter is less than 2 microns;
  10. Password protection setting parameters;
  11. The measurement data U disk is saved in EXCEL format for easy editing and processing;
  12. Can be upgraded to computer control system;
  13. Force calibration accessories are included for test force calibration of various dynamometers.

II、Main technical parameters:

Product model TMHR-150SX
Rockwell initial test force 10Kgf (98.07N)
Rockwell total test force 60kgf(588.4N), 100kgf(980.7N), 150kgf(1471N)
Hardness scale HRA, HRB, HRC, HRD, HRE, HRF, HRG, HRH, HRK
Hardness indicator measurement resolution 0.01 HR
Maximum permissible deviation of initial test force “Plus or minus 0.2%
Maximum allowable deviation of total test force “Plus or minus 0.1%
Depth measuring system resolution 0.01 mu m
Maximum allowable error for the depth measurement system “Plus or minus 0.2 microns
Standard uncertainty < 0.2HR
Ball indenter diameter error (mm) + / – 0.002
enforce standards JJG112; ASTM E18
Retention time 0-60s
Hardness read LCD display
Maximum height of specimen (mm) 20
Data output Flash drive
Power supply AC220V
Host weight About 20kg

III、Accessories:

Name Quantity Name Quantity
Standard diamond indenter

(US import)

1 Standard ball indenter

Carbide, diameter 1.5875mm

1
Flash drive 1 Force calibration attachment 1 set