TMHR-150SX Rockwell hardness standard machine
I、Product Introduction
TMHR-150SX Rockwell hardness standard machine is a new generation of Rockwell hardness standard machine with international advanced level; Mainly suitable for the use of standard Rockwell hardness block verification, Rockwell hardness measurement value transfer.
The standard machine has the characteristics of high automation, high precision, good stability, etc., and fully meets the index requirements of the Rockwell hardness standard machine in China and the United States.
- Rockwell A ~ K scale;
- Automatic closed-loop load control, one-key operation, automatic loading, automatic unloading and indentation depth measurement;
- Automatic application of initial test force and main test force, automatic removal of main test force;
- The maximum allowable deviation of the initial test force is less than ±0.2%, and the maximum allowable deviation of the total test force is less than ±0.1%;
- The resolution of the indentation depth measurement system is 0.1μm, and the maximum allowable error is less than ±0.2μm;
- It can be set and controlled strictly during each period of time during loading and unloading.
- The uncertainty of test time is less than 0.5 seconds;
- The Angle error of diamond compression head Angle is less than ± 0.1
- Diameter error of ball indenter is less than 2 microns;
- Password protection setting parameters;
- The measurement data U disk is saved in EXCEL format for easy editing and processing;
- Can be upgraded to computer control system;
- Force calibration accessories are included for test force calibration of various dynamometers.
II、Main technical parameters:
Product model | TMHR-150SX |
Rockwell initial test force | 10Kgf (98.07N) |
Rockwell total test force | 60kgf(588.4N), 100kgf(980.7N), 150kgf(1471N) |
Hardness scale | HRA, HRB, HRC, HRD, HRE, HRF, HRG, HRH, HRK |
Hardness indicator measurement resolution | 0.01 HR |
Maximum permissible deviation of initial test force | “Plus or minus 0.2% |
Maximum allowable deviation of total test force | “Plus or minus 0.1% |
Depth measuring system resolution | 0.01 mu m |
Maximum allowable error for the depth measurement system | “Plus or minus 0.2 microns |
Standard uncertainty | < 0.2HR |
Ball indenter diameter error (mm) | + / – 0.002 |
enforce standards | JJG112; ASTM E18 |
Retention time | 0-60s |
Hardness read | LCD display |
Maximum height of specimen (mm) | 20 |
Data output | Flash drive |
Power supply | AC220V |
Host weight | About 20kg |
III、Accessories:
Name | Quantity | Name | Quantity |
Standard diamond indenter
(US import) |
1 | Standard ball indenter
Carbide, diameter 1.5875mm |
1 |
Flash drive | 1 | Force calibration attachment | 1 set |
IV、Options:
1.Computer processing system for hardness data
2.Surface Rockwell hardness standard machine accessories
Write your message here and send it to us